Rise time is considered only when calibrating a pulse waveform (or system), because of the diffilculty of technique, but the other one main parameter-overshoot is not be calibrated Relationship of the overshoot between pulse waveform, measurement system and response waveform is analyzed fully in this paper, and the calibration procedure and formula for overshoot are given. It is important that the method can estimate overshoot parameter with higher precision anti without the bandwidth restriction. It is valuable for overshoot calibration in practice.
Keyword:
Reprint Author's Address:
Email:
Source :
ICEMI'2003: PROCEEDINGS OF THE SIXTH INTERNATIONAL CONFERENCE ON ELECTRONIC MEASUREMENT & INSTRUMENTS, VOLS 1-3
Year: 2003
Page: 1228-1232