Abstract:
An automatic method is proposed to solve the registration problem, which aligns a single 2D fluoroscopic image to a 3D image volume without demanding any additional media like calibration plate or user interactions. First, a mathematic projection model is designed which can reduce the influence of projection distortion on parameter optimization and improve the registration accuracy. Then, a two stage optimization method is proposed, which enables a robust registration in a wide parameter space. Furthermore, an automatic registration framework is proposed based on the Fourier-Mellin robust image comparison descriptor. Experimental results show that the registration method has a high accuracy with average rotation error of 0.6 degree and average translation error of 1.4 mm. Copy Right by Hight Technology Letters Press.
Keyword:
Reprint Author's Address:
Email:
Source :
High Technology Letters
ISSN: 1006-6748
Year: 2013
Issue: 4
Volume: 19
Page: 378-383
Cited Count:
WoS CC Cited Count: 0
SCOPUS Cited Count:
ESI Highly Cited Papers on the List: 0 Unfold All
WanFang Cited Count:
Chinese Cited Count:
30 Days PV: 1
Affiliated Colleges: