Abstract:

Objective: To clarify the influence from left ventricular assist device (LVAD) cannula anastomosis on micro-emboli movement by numerical simulation methods. Methods: Firstly, a patient-specific aorta geometric model was reconstructed based on CT scan images of a patient with heart failure. Three geometric models of LVAD support were established according to clinical practice, and the outflow cannula of LVAD was anastomosed at the three different locations of ascending aorta (the anterior wall, between anterior wall and lateral wall, the lateral wall of ascending aorta). Secondly, the computational fluid dynamic (CFD) approach combined with the discrete phase theory was used to calculate the hemodynamic states, and Fluent finite element software was used for numerical simulation of the fluid. Finally, the micro-emboli distribution flow pattern, wall shear stress (WSS) and velocity vector were extracted to evaluate the hemodynamic state. Results: The probability of micro-emboli entering the brain vessel at the three different anastomosis locations were 23.6%, 33.8% and 36.7%, respectively. Besides, vortex was observed around the anastomosis locations. Conclusions: The different anastomosis locations of LVAD cannula can significantly change the hemodynamic environment, which will influence micro-emboli movement in the aorta. The research findings provide some insights and theoretical support for optimizing the anastomosis to decrease occurrence risk of stroke in clinic. Copyright © 2016 by the Editorial Board of Journal of Medical Biomechanics.

Keyword:

Artificial heart bump; Hemodynamics; Micro-emboli distribution; Stroke

Author Community:

  • [ 1 ] [Zhang, Y.-G.]Biomedical Engineering Center, Beijing University of Technology, Beijing, 100124, China
  • [ 2 ] [Gao, B.]Biomedical Engineering Center, Beijing University of Technology, Beijing, 100124, China
  • [ 3 ] [Chang, Y.]Biomedical Engineering Center, Beijing University of Technology, Beijing, 100124, China

Reprint Author's Address:

  • 常宇

    [Chang, Y.]Biomedical Engineering Center, Beijing University of TechnologyChina

Show more details

Related Keywords:

Related Article:

Source :

Journal of Medical Biomechanics

ISSN: 1004-7220

Year: 2016

Issue: 3

Volume: 31

Page: 199-205

Cited Count:

WoS CC Cited Count: 0

SCOPUS Cited Count:

ESI Highly Cited Papers on the List: 0 Unfold All

WanFang Cited Count:

Chinese Cited Count:

30 Days PV: 7

Online/Total:501/10564081
Address:BJUT Library(100 Pingleyuan,Chaoyang District,Beijing 100124, China Post Code:100124) Contact Us:010-67392185
Copyright:BJUT Library Technical Support:Beijing Aegean Software Co., Ltd.