Abstract:
This paper constructed a calibration system of oscilloscope based on the optical pulse source. The 70GHz sampling module 86118A of Agilent 86100C was calibrated based on optical pulse source using the standard pulse method. In the calibration process, the influences on calibration results of three kinds of time base errors were researched for the first time, e. g., time base drift was removed using a cross-correlation algorithm, time base distortion was corrected using the least squares method, and jitter was removed by using PDF deconvolution. Experiment results showed that calibration of 70GHz broadband oscilloscope was realized, the time base errors could be effectively removed by this method, and the accuracy of the calibration was satisfied. © 2018, Chinese Institute of Electronics. All right reserved.