Abstract:

In this paper is introduced a design of two-dimensional dynamic test system for mechanical performance of biosoft tissue. The external force exterted on the tissue and deformation of the tissue are measured to reflect dynamic properties of soft tissue. Experiments were carried out using rabbit abdomen as the object. This system was used to measure hysteresis, relaxation and creeping characteristics. The feasibility of the system was proved.

Keyword:

Biomechanics; Constitutive equation; Creep; Hysteresis; Relaxation; Soft-tissue

Author Community:

  • [ 1 ] [Hao, D.]Beijing Polytechnic University, Beijing 100022, China
  • [ 2 ] [Liu, Z.]Beijing Polytechnic University, Beijing 100022, China
  • [ 3 ] [Li, H.]Beijing Polytechnic University, Beijing 100022, China