Indexed by:
Abstract:
本文构建了基于光脉冲源的示波器校准系统,用标准脉冲法实现了对宽带取样示波器Agilent 86100C的70GHz取样模块86118A的校准.在校准实现过程中,首次综合研究了三种时基误差对校准结果的影响,并采用互相关算法去除时基漂移,最小二乘法修正时基失真,利用反卷积去除时基抖动.实验结果表明,本文方法实现了对70GHz宽带取样示波器的校准,有效地去除了时基误差,满足校准精度的要求.
Keyword:
Reprint Author's Address:
Email:
Source :
电子学报
Year: 2018
Issue: 04
Volume: 46
Page: 945-951
Cited Count:
WoS CC Cited Count: 0
SCOPUS Cited Count:
ESI Highly Cited Papers on the List: 0 Unfold All
WanFang Cited Count:
Chinese Cited Count:
30 Days PV: 3
Affiliated Colleges: