通过理化检验和宏微观检查,分析了脉动真空灭菌器内腔开裂失效的原因.灭菌器在制作过程中的加强筋板的焊接造成了对内腔的残余拉应力,灭菌器T作介质高温蒸汽中含有微量的Cl-.在残余拉应力、高温以及Cr-的共同作用下,引发脉动真空灭菌器内腔表面钝化膜破裂并导致了应力腐蚀开裂.为解决脉动真空灭菌器内腔开裂问题提出了一些可行性建议.